Sykes, Jonathan (2010) Quantification of geometric uncertainties in image guided radiotherapy. PhD thesis, University of Leeds.
Abstract
The aim of this thesis is to determine if the geometric uncertainties that are introduced into the image guided radiotherapy (IGRT) process by Cone Beam CT (CBCT) based IGRT equipment are sufficiently small that they do not pose a significant risk of geometrical error in treatment delivery. This was performed by quantifying and
investigating the geometric uncertainties introduced by; (1) calibration of the image geometry, (2) correction of patient position performed by automatic treatment couch
systems and (3) automatic image registration of the localisation image with a reference image. In addition, the feasibility of providing user feedback on the likelihood of
accurate image registration was investigated. A method was developed using supervised machine learning based on the shape of the image registration algorithm's similarity metric surface.
The geometric uncertainties introduced by image calibration and couch positioning were both shown to be less than 1 mm and therefore do not contribute significantly to
the overall uncertainties in the IGRT process. Image registration performance for image guidance based on the bony anatomy of the skull was shown to be reproducible, accurate and robust with errors typically less than 1 mm. Moreover, image registration performance did not deteriorate significantly as imaging dose was reduced. For image guidance based on the soft tissues of the prostate, image registration performance was satisfactory for some CBCT images resulting in errors less than 2 mm. However, with the majority of CBCT images, image registration was highly irreproducible with high frequencies of failure. The user feedback of image registration quality was able to
correctly classify 84% of image registrations into categories of good, acceptable and unacceptable. No unacceptable classifications were classed as good.
CBCT based IGRT equipment does not introduce significant risks into the IGRT process however, appropriate quality assurance measures should be implemented to safeguard against equipment failure and drift since previous system calibration. Automatic image registration of the soft-tissues of the prostate cannot be relied upon for
clinical use and therefore it should be used in conjunction with manual methods.
Metadata
Supervisors: | Magee, D. and Brettle, D. and Thwaites, D. |
---|---|
ISBN: | 978-0-85731-066-8 |
Awarding institution: | University of Leeds |
Academic Units: | The University of Leeds > Faculty of Medicine and Health (Leeds) > School of Medicine (Leeds) |
Identification Number/EthosID: | uk.bl.ethos.570096 |
Depositing User: | Repository Administrator |
Date Deposited: | 15 Apr 2013 12:23 |
Last Modified: | 07 Mar 2014 11:24 |
Open Archives Initiative ID (OAI ID): | oai:etheses.whiterose.ac.uk:3864 |
Download
JRSykes_Thesis_Final_150211_200238858
Filename: JRSykes_Thesis_Final_150211_200238858.pdf
Licence:
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 2.5 License
Export
Statistics
You do not need to contact us to get a copy of this thesis. Please use the 'Download' link(s) above to get a copy.
You can contact us about this thesis. If you need to make a general enquiry, please see the Contact us page.