Hua, Xijin (2013) Contact mechanics of metal on polyethylene hip replacements. PhD thesis, University of Leeds.
Abstract
Metal-on-ultra high molecular weight polyethylene (UHMWPE) total hip replacement (THR) has been the most popular and clinically successful hip prosthesis to date. The long-term performance of THR depends on both the tribological characteristics and biomechanical behaviour of the prosthesis. This project focused on understanding the contact mechanics and mechanical behaviour of cemented and cementless metal-on-UHMWPE THRs under different conditions using a computational approach.
Three-dimensional (3D) computational models of THRs with realistic pelvic bone were developed. Two typical bearings, the Charnley hip and the Pinnacle cup system, were investigated. The effect of different factors on the contact mechanics and cement stresses for Charnley THR were examined. Additionally, the contact mechanics and mechanical behaviour of Pinnacle THR under daily activities, standard and microseparation conditions were analysed.
The cup angles and penetration depths in the cup, and the sizes of the components were found to have a significant effect on the contact mechanics and cement stresses for Charnley THR. The stresses at the bone-cement interface for the Charnley THR with outer diameter of 40 mm were predicted to be higher than that of 43 mm, the difference was found to be consistent with the clinical observation of different aseptic loosening rates.
The cup angles and radial clearances were found to have a synergistic effect on the contact mechanics of Pinnacle THR. Edge loading on both articulating surface and backside surface of the liner was observed during some daily activities due to steep cup inclination angles and smaller radial clearance. The introduction of microseparation into the gait cycle, especially when combined with steep cup inclination angles, resulted in concentrated stresses and plastic deformation in the liner, which would cause potential damage to the liner. Therefore, it is critically important to reduce the levels of rotational and translational mal-positioning of the components clinically.
Metadata
Supervisors: | Fisher, John and Jin, Zhongmin and Wilcox, Ruth and Wang, Ling |
---|---|
ISBN: | 978-0-85731-560-1 |
Awarding institution: | University of Leeds |
Academic Units: | The University of Leeds > Faculty of Engineering (Leeds) > School of Mechanical Engineering (Leeds) |
Identification Number/EthosID: | uk.bl.ethos.595227 |
Depositing User: | Repository Administrator |
Date Deposited: | 26 Mar 2014 16:26 |
Last Modified: | 03 Sep 2014 10:49 |
Open Archives Initiative ID (OAI ID): | oai:etheses.whiterose.ac.uk:5568 |
Download
Final eThesis - complete (pdf)
Filename: PhD thesis_Xijin Hua_200479774_Hardcopy.pdf
Licence:
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 2.5 License
Export
Statistics
You do not need to contact us to get a copy of this thesis. Please use the 'Download' link(s) above to get a copy.
You can contact us about this thesis. If you need to make a general enquiry, please see the Contact us page.